Visualize a new path forward
Changing the course of how diseases are treated and ultimately cured requires a comprehensive understanding of complex cellular phenotypes and their interrelationships in the spatial context of the tissue microenvironment. To meet this great challenge, we developed the Hyperion™ Imaging System.
The Hyperion Imaging System brings proven CyTOF® technology together with imaging capability to empower simultaneous interrogation of 4 to 37 protein markers using Imaging Mass Cytometry™ (IMC™). Using this system, you can deeply interrogate tissues and tumors at subcellular resolution while preserving the information in tissue architecture and cellular morphology to uncover new biomarker correlations and cell interactions. With the ability to utilize up to 135 channels to detect additional parameters, the Hyperion Imaging System is ideal to meet your image analysis needs today and well into the future.
Comprehensive Highly multiplexed immunohistochemistry (IHC) using proven CyTOF technology enables simultaneous analysis of 4 to 37 protein markers from a single scan |
Contextual Identify new protein biomarkers and deeply profile tissue microenvironments at subcellular resolution while preserving the information in tissue architecture and cellular morphology |
Powerful Preserve precious samples and reduce variability by eliminating dependency on serial sections |
Hear what researchers are saying about Imaging Mass Cytometry and the Hyperion Imaging System:

Learn more
Complete the form below to stay up to date on new Hyperion™ Imaging System antibodies, protocols and other developments.
Technology
Imaging Mass Cytometry (IMC) on the Hyperion Imaging System uniquely combines a precisely directed laser beam focused at 1 μm to collect biological samples stained with metal-tagged Maxpar® antibodies and directs these tags for analysis with proven CyTOF technology. Bringing together the high-multiplex capabilities of mass cytometry with imaging using metal-tagged antibodies allows simultaneous interrogation of 4 to 37 protein markers in tissues and tumors at subcellular resolution while preserving the information in tissue architecture and cell morphology. Now researchers can detect discrete signals from each ionized metal tag based on differences in mass instead of wavelength at 1 Da resolution, and in contrast to fluorescence results, get uniform staining with minimal background.
Workflow
Support
components
IFCs, Assays and Reagents
Software
Specifications
Hyperion Imaging System
Channels | 135 |
---|---|
Mass range | 75–209 amu |
Abundance sensitivity | 0.3% for 159Tb |
Frequency | 200 pixels/sec |
Detection limit | ≥400 copies per μm2 |
Dynamic range | 4 orders of magnitude |
Calibration | Automated |
Operating system | Windows® 7 Pro 64-bit |
Data storage | >6 TB RAID (mirrored) |
Switch time (between modes) | 12 hr |
Cross-cell contamination (redeposition) | ≤2% (selected direction) |
Crosstalk pixel to pixel | ≤15% at 200 pixels/sec |
Wet tissue thickness for full ablation | ≤7 μm |
Addressable sample size on slide | ≥15 mm x 45 mm |
Optical view of sample | ≥ 250 μm x 250 μm |
File type | TXT, multipage TIFF, OME-TIFF, MCD |
Scan area | ≥1 mm2/2 hr (@200 Hz) |
Workstation
Motherboard | HP® Z840 Workstation |
---|---|
CPU 1 and 2 | Intel® Xeon® E5-2643 v4, 3.40 GHz, 2400 6C |
RAM | 64 GB DDR4-2400 (8 x 8 GB) 2 CPU RegRAM |
Graphics card | NVIDIA® Quadro® K620 2 GB Graphics (DL-DVI+DP) |
DVD | 9.5 mm Slim SuperMulti DVD-RW 1st ODD |
Keyboard | HP USB keyboard US |
Mouse | HP USB optical mouse |
Chassis | HP Z840 1,125 W (1,450 W/200 V) 90% efficient chassis |
RAID controller | LSI 9270-8i SAS 6Gb/s ROC RAID Card |
Storage drives | HP 2 x 256 GB SATA SSD RAID 1
Western Digital® 4 x 4 TB 7200 RPM SATA, RAID 10 |
Monitor
Monitor | LG® 34" Class 21:9 UltraWide® LED Monitor |
---|---|
Screen size | 34.1 in (86.6 cm) |
Data File Size*
*1,000 x 1,000 μm region of interest (ROI)
MCD | 4-5 MB/channel |
---|---|
TXT | 6-8 MB/channel |
Hyperion Tissue Imager Dimensions*
*The Hyperion Tissue Imager must be placed directly in front of the Helios instrument.
Width | 56 cm (22 in) |
---|---|
Height | 134 cm (53 in) |
Depth | 56 cm (22 in) |
Weight | 159 kg (350 lb) |
Helios Dimensions
Width | 103 cm (41 in) |
---|---|
Height | 132 cm (52 in) |
Depth | 87 cm (35 in) |
Weight | 320 kg (705 lb) |
Chiller Size
Width | 38 cm (15 in) |
---|---|
Height | 64 cm (25 in) |
Depth | 67 cm (27 in) |
Weight | 81 kg (178 lb) |
Monitor Size
Width | 83.05 cm (32.70 in) |
---|---|
Height | 54.86 cm (21.60 in) |
Depth | 23.87 cm (9.40 in) |
Weight | 8.40 kg (18.50 lb) with stand |